OI-LLT100-EN Rev. D |
User Guide
45
Table 4 Errors
#
Error message
Action
0
Level Sensor Failure
Contact ABB service for replacement.
1
Temperature Sensor Failure
Contact ABB service for replacement.
2
Illegal Memory Access
Reset device. Contact ABB if condition persists.
3
Safety Function Flow Failure
Reset device. Contact ABB if condition persists.
4
Level Sensor Out of Limits
Check for level sensor failure.
5
Temperature Sensor Out of Limits
Check for temperature sensor failure.
6
Self-Test Failure. Reset device
Contact ABB if condition persists.
10
Level Sensor. Out of Range
Check process.
16
Sensor board failure
Contact ABB service for replacement.
17
Sensor Board Memory Failure
Reset device. Contact ABB if condition persists.
18
Sensor Board NV Write Error
Reset device. Contact ABB if condition persists.
19
Device Reset Required
Cycle power for changes to take effect.
20
Sensor Communication Failure
Reset device. Contact ABB if condition persists.
23
Electronics NV Failure
Contact ABB service.
25
Electronics NV Syncing Information
Contact ABB service.
30
HMI Validation Test Failure Reset device
Replace LCD if condition persists.
31
Sensor Window Obstructed/Dirty
Clean window. Contact ABB if condition persists.
32
Electronics Insuff Input Voltage
Insufficient voltage to operate device.
33
Amb. Temp Out of Range
Readings may be invalid due to improper ambient temperature. Correct ambient temperature.
36
Data Simulation Warning
One or more process values are in SIM mode.
37
Alarm Simulation Warning
One or more alarms are in simulation mode.
38
Current Output in Fixed Mode
Process should be in manual control mode.
39
Primary Current Uncertain
Contact ABB support if condition persists.
40
Electronics ROM Failure
Contact ABB support if condition persists.
41
Electronics RAM Failure
Contact ABB support if condition persists.
42
Target lost Error
Check Process conditions.
43
Primary Current Saturated
Contact ABB support if condition persists.
44
Current Output Failure
Contact ABB support if condition persists.
45
Process Media Warning
Monitor Process Conditions.
46
Process Media Alarm
Monitor Process Conditions.
47
Invalid algorithm parameter
Adjust configured Parameters.
Summary of Contents for LLT100
Page 8: ...This page intentionally left blank ...
Page 10: ...This page intentionally left blank ...
Page 18: ...This page intentionally left blank ...
Page 22: ...This page intentionally left blank ...
Page 26: ...This page intentionally left blank ...
Page 44: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 66: ...This page intentionally left blank ...
Page 67: ......