54
User Guide
| OI-LLT100-EN Rev. D
LLT100
Laser level transmitter
Tube model
Ambient temperature (°C)
Air flow needed?
(minimum recommended flow, if needed)
Air pressure at cooling tube
air inlet*
P920
20
30
40
50
Max. pr
ocess
temperatur
e
(°
C
)
205
155
110
65
No
–
300
225
150
75
Yes (1 scfm)
0.50 psi (0.05 bar)
280
185
85
Yes (2 scfm)
1 psi (0.075 bar)
350
230
105
Yes (3 scfm)
2.5 psi (0.2 bar)
305
130
Yes (4 scfm)
5 psi (0.35 bar)
165
Yes (5 scfm)
8.5 psi (0.6 bar)
Tube model
Ambient temperature (°C)
Air flow needed?
(minimum recommended flow, if needed)
Air pressure at cooling tube
air inlet*
P921
20
30
40
50
Max. pr
ocess
temperatur
e
(°
C
)
175
85
No
–
150
Yes (1 scfm)
0.50 psi (0.05 bar)
180
Yes (2 scfm)
1.25 psi (0.1 bar)
Yes (3 scfm)
2.75 psi (0.2 bar)
Yes (4 scfm)
6.25 psi (0.45 bar)
Tube model
Ambient temperature (°C)
Air flow needed?
(minimum recommended flow, if needed)
Air pressure at cooling tube
air inlet*
P922,
P923, P924
20
30
40
50
Max. pr
ocess
temperatur
e.
(°
C
)
200
90
No
–
155
Yes (1 scfm)
0.50 psi (0.05 bar)
280
200
Yes (2 scfm)
1.25 psi (0.1 bar)
270
Yes (3 scfm)
2.75 psi (0.2 bar)
Yes (4 scfm)
6.25 psi (0.45 bar)
Summary of Contents for LLT100
Page 8: ...This page intentionally left blank ...
Page 10: ...This page intentionally left blank ...
Page 18: ...This page intentionally left blank ...
Page 22: ...This page intentionally left blank ...
Page 26: ...This page intentionally left blank ...
Page 44: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 66: ...This page intentionally left blank ...
Page 67: ......