Menu: Calibrate (continued)
Level Offset
. . .
Interface Offset
Temperature
Offset
Menu/Parameter
Value Range
Description
Calibrate/Level Calibration
Calibration Points
‘Out’ values range must be within 5% of the ‘In’
values range
SVL – Sensor value at Level
Points – Corresponds to points 00 and 01
In – Sensor value
Out – Level value
Reset Calibration
Resets calibration points to factory defaults
Calibrate/Interface Calibration
Calibration Points
‘Out’ values range must be within 5% of the ‘In’
values range
SVI – Sensor value at Interface
Points – Corresponds to points 00 and 01
In – Sensor value
Out – Level value
Reset Calibration
Resets calibration points to factory defaults
Calibrate/Process Temp Calibration
Temperature Offset
Reset Calibration
Resets calibration to factory defaults