LMT10 0 A N D LMT 20 0 |
M AG N E TOS T R I C T I V E L E V EL T R A N SM I T T ER S | FF O I/L M T 10 0/ 2 0 0/FF - EN R E V. A
29
Easy Setup
This menu allows the verification and the setting of parameters
for the basic configuration of the LMT Series level transmitters
Easy Setup guides the user to the choice of interface language,
the tag number configuration, the engineering units, the upper
range value and the lower range value (URV and LRV) and the
display visualization mode (the value that needs to be
visualized on the HMI)
Device Setup
This menu allows the verification and the establishing of
parameters related to the whole LMT Series of devices This
includes enabling write-protection, setting passwords for
access control, setting process variables (unit, LRV and URV),
selecting transfer functions (linearization type and low flow
cutoff) and scaling output (unit according to the measurement
and LRV / URV) The last sub-menu allows the user to reset all
the parameters to the default configuration
Display
The Display menu allows the setup of functions relevant to the
display itself These include the display language and contrast
It is also possible to customize the screen display to one or two
lines with or without the bar graphs There is the possibility of
setting a protection password (security) and the display scaling
(distance, linearization and temperature format)
Calibrate
The Calibrate menu allows the local calibration of the
instrument This includes setting calibration points and the
offset value for the level, liquid interface (for two liquids) and
temperature
Diagnostics
The Diagnostics menu allows the user to monitor diagnostic
messages that relate to the waveform, signal polarity,
simulation and history information
Device Info
The Device Info menu enables the user to retrieve all device
information The menu-driven structure shows the user the
sensor type, the hardware and software revisions, the high and
low sensor limits, and the minimum applicable span
Communication
The last section of the structured menu items, the
Communication menu allows the user to change the
communication tag and the MULTI-DROP mode with FF address
numbers for the device It is also where variables are assigned
to FF addresses (PV, SV, TV and QV)