30
LMT S E R I E S |
M AG N E TOS T R I C T I V E L E V EL T R A N SM I T T ER | O I/L M T 10 0/ 2 0 0 - EN R E V. B
Switching to operator menu
The operator menu can be used to display diagnostics
information and select which operator pages to display
1
Press
to switch to the operator menu
2
Press
or
to select a submenu
3
Press
to confirm your selection
Menu
Description
/ Operator Menu
Error / Alarm of the electronics
Diagnostics
Error / Alarm of the sensor
Operator Page 1
Select the “Diagnostics submenu” See section 8
“Diagnostic History”
Operator Page 2
Select the operator page to displayed
Operator Page 3
Select the operator page to displayed
Operator Page 4
Select the operator page to displayed
Signals View
Error / Alarm due to the present operating
conditions
Area
Description
Electronics
Error / Alarm of the electronics
Sensor
Error / Alarm of the sensor
Status
Alarm due to the present device status
Configuration
Error / Alarm due to the present operating
conditions
Invoking the error description
In case of an error, a message consisting of an icon and text
appears at the bottom of the process display The displayed
text indicates where the error has occurred
1
Press
to switch to the operator menu
2
Press
or
to navigate to select a submenu
3
Press
to confirm your selection
The first line indicates where the error occurred
The second line shows the unique id
The next lines give a brief description of the error and Its
remedy
Switching to the configuration level parameter entry
The device parameters can be displayed and changed on the
configuration level
1
Press
to switch to the configuration menu
2
Press
or
to select an access level
3
Press
to confirm your selection
IMPORTANT NOTE
There are four access levels as follows:
– “Read Only” level disables all entries No parameter can
be modified
– “Standard” level can edit some parameters
– “Advanced” level can edit all parameters
– “Service” level is reserved for ABB technician access
Passwords can be defined for the “Standard” and
“Advanced” levels
Document your password so that it can be retrieved
later
4
Enter the corresponding password
5
Press
to switch to the information level The HMI display
now indicates the first menu item at the configuration level
6
Press
or
to select a menu
7
Press
to confirm your selection