56
LMT S E R I E S |
M AG N E TOS T R I C T I V E L E V EL T R A N SM I T T ER | O I/L M T 10 0/ 2 0 0 - EN R E V. B
STEP 3:
Press
or
to navigate to the Diagnostic menu
Press “Select”
to enter the menu
STEP 4:
Press
or
to get to “Waveform” menu
Press “Select”
to enter the menu
STEP 5:
Press
or
to get to the desired position of the
waveform
Press “Select”
to enter the waveform screen
IMPORTANT NOTE
• At Sensor Ref Pt displays the signal starting from the
beginning of the probe, which is the same as from the
enclosure down
• “At Level” displays the signal with the Level position centered
in the screen, unless other limit factors apply but in any case
the level position should be visible in the screen
• “At Distance” displays the signal starting from the distance
specified by the user The Level position must be centered in
the screen, unless other limit factors apply but in any case the
level position should be visible in the screen
• “End of Probe” displays the signal the tip of the probe
Review or Edit device measurement parameters
THRESHOLD:
Press
to scroll the cursor to the device measurement
parameters
Press
or
keys to select THD (threshold)
Press
to edit the threshold value
Press
to scroll from one digit to another
Press
or
keys to edit each digit
Press
to complete this operation and OK/accept the input
value
PULSE WIDTH:
Press
and
keys to select PLS (pulse width)
Press
to edit the PLS value
Press
to scroll from one digit to another
Press
or
keys to edit each digit
Press
to complete this operation and OK/accept the input
value
BLANKING / BLOCKING DISTANCE:
Press
or
keys to select BLK (Blanking Distance)
Press
to edit the BLK value
Press
to scroll from one digit to another
Press
or
keys to edit each digit
Press
to complete this operation and OK/accept the input
value