Error Message
Tx LCD Message
Possible Cause
Suggested Action
Tx Response
S224.043
Primary Current
Saturated
Current saturated high or low
Check process conditions and/or device
configuration The compatibility of level
transmitter model and process conditions has to
be checked Probably a different transmitter type
is required
No effect
_156.025
Electronics NV
Syncing
Write non-volatile command
received from CB
Information only
No effect
Non-volatile Storage in
progress
C154.047
Sensor Config
Warning
The Level sensor trim is set to
a default value
Contact service to perform a factory level sensor
trim
No effect
The Temperature trimming
configuration is corrupted
C200.019
Device Reset
Required
Device change requires restart
Restart device
No effect
C211.018
Sensor Board NV
Write Error
An attempt to save data to the
sensor module has failed
Restart the device If the condition persists
contact service
No effect
F098.001
Temperature Sensor
Failure
A temperature measurement
has resulted in a value outside
of the operating range
Restart the device If the condition persists
contact service for a replacement
Analog signal to alarm
F194.031
Sensor Board Fault
An error occurred in the sensor
module Possible causes could
be a failed component in the
sensor electronics, a
configuration error, or the
sensor assembly
Using a HART configurator (DTM - Hand held) or
the local HMI, navigate to the Diagnostics ->
Waveform display and confirm the quality of the
signal The signal peak should visibly cross the
threshold line and should measure approximately
2 Volts. Contact ABB service for additional
support or a replacement if the condition
persists
Analog signal to alarm
F196.020
Sensor
Communication
Failure
An error occurred in the
communication between the
sensor and electronics
modules
Restart the device If the condition persists
contact service for a replacement
Analog signal to alarm
F198.000
Level Sensor Failure
An error occurred in the sensor
module Possible causes could
be a failed component in the
sensor electronics or an error
changing the configuration
Restart the device If the condition persists
contact service for a replacement
Analog signal to alarm
F212.017
Sensor Board
Memory Failure
Writing to the sensor non-
volatile memory was not
successful
The sensor should be replaced as soon as
possible
Analog signal to alarm