50
L ST 3 0 0 |
CO M PAC T U LT R A S O N I C L E V EL T R A N SM I T T ER | L S T 3 0 0 – EN R E V. E
Dry calibration procedure
1
Log on to the LST300 at the
Standard
or
Advanced
level .
2
Select
Calibrate
from the main menu .
3
Select
Calibration
from the “Calibrate” submenu .
4
Select
Dry
from the “Calibration” submenu .
5
Configure the four parameters in this menu using the data
already obtained earlier in section “Dry calibration” .
The level dry calibration is immediately active on the LST300 .
Wet calibration
Wet calibration can be done in position, and the measurement
is performed directly from the calibration menu . No
measurement is required prior to calibration . It is however
required that the level can be adjusted to the calibration
positions while performing the calibration .
Wet calibration procedure
1
Follow the steps 1 to 3 in section “Dry calibration procedure”
to navigate to the “Calibration” submenu .
2
Select
Wet
from the “Calibration” submenu .
3
Ensure the LST300 is measuring the upper calibration point,
by ensuring the conditions as indicated in figure 26 “Upper
calibration point measured” .
Figure 30 Upper calibration point measured
4
Configure the
Upper Level Point
as the known value of the
level at a position close to the top of the tank .
Summary of Contents for LST300 Series
Page 26: ...26 LST300 COMPACT ULTR ASONIC LEVEL TR ANSMITTER LST300 EN REV E...
Page 27: ...LST300 COMPACT ULTR ASONIC LEVEL TR ANSMITTER LST300 EN REV E 27...
Page 28: ...28 LST300 COMPACT ULTR ASONIC LEVEL TR ANSMITTER LST300 EN REV E...
Page 59: ...LST300 COMPACT ULTR ASONIC LEVEL TR ANSMITTER LST300 EN REV E 59 Note...