Navigator 540 | Transmitter |
COM/AWT540/PB–EN Rev. C
47
43012
Sensor 2 – Concentration
units
7 – ppb
8 – ug/l
9 – ug/Kg
43013
Sensor 2 – Filter type
0 – Off
1 – Minimum value
2 – Maximum value
3 – Average value
4 – Sliding average
43014
Sensor 2 – Filter time
5 to 100 Samples / Seconds
43015
Sensor 2 – Wetting routine
procedure
0 – Disabled
1 – Enabled
43016
Sensor 2 – Flow measurement
0 – Disabled
1 – Enabled
43017
Sensor 2 – Out of sample
detection
0 – Disabled
1 – Enabled
43018
Sensor 2 – Scheduled
calibration type
0 – Scheduled calibration
disabled
1 – One point calibration
(low standard)
2 – One point calibration
(high standard)
3 – Two point calibration
4 – Regene two point
calibration
5 – One point air calibration
43019
Sensor 2 – Scheduled
calibration frequency units
0 – Frequency off
1 – Daily
2 – Weekly
3 – Monthly
43020
Sensor 2 – Scheduled
calibration frequency daily
0 – Daily
1 – Two days
2 – Three days
3 – Four days
4 – Five days
5 – Six days
6 – Seven days
43021
Sensor 2 – Scheduled
calibration frequency weekly
0 – Weekly
1 – Two weeks
2 – Three weeks
3 – Four weeks
4 – Five weeks
5 – Six weeks
6 – Seven weeks
7 – Eight weeks
43022
Sensor 2 – Scheduled
calibration frequency monthly
0 – Monthly
1 – Two months
2 – Three months
3 – Four months
4 – Six months
5 – Twelve months
43023
Sensor 3 – Concentration
units
7 – ppb
8 – ug/l
9 – ug/Kg
Modbus
address
Description
Valid range
Table B.21: Single-stream sensors – Units, filter and
scheduled calibration configuration (continued)
43024
Sensor 3 – Filter type
0 – Off
1 – Minimum value
2 – Maximum value
3 – Average value
4 – Sliding average
43025
Sensor 3 – Filter time
5 to 100 Samples / Seconds
43026
Sensor 3 – Wetting routine
procedure
0 – Disabled
1 – Enabled
43027
Sensor 3 – Flow measurement
0 – Disabled
1 – Enabled
43028
Sensor 3 – Out of sample
detection
0 – Disabled
1 – Enabled
43029
Sensor 3 – Scheduled
calibration type
0 – Scheduled calibration
disabled
1 – One point calibration
(low standard)
2 – One point calibration
(high standard)
3 – Two point calibration
4 – Regene two point
calibration
5 – One point air calibration
43030
Sensor 3 – Scheduled
calibration frequency units
0 – Frequency off
1 – Daily
2 – Weekly
3 – Monthly
43031
Sensor 3 – Scheduled
calibration frequency daily
0 – Daily
1 – Two days
2 – Three days
3 – Four days
4 – Five days
5 – Six days
6 – Seven days
43032
Sensor 3 – Scheduled
calibration frequency weekly
0 – Weekly
1 – Two weeks
2 – Three weeks
3 – Four weeks
4 – Five weeks
5 – Six weeks
6 – Seven weeks
7 – Eight weeks
43033
Sensor 3 – Scheduled
calibration frequency monthly
0 – Monthly
1 – Two months
2 – Three months
3 – Four months
4 – Six months
5 – Twelve months
43034
Sensor 4 – Concentration units
7 – ppb
8 – ug/l
9 – ug/Kg
Modbus
address
Description
Valid range
Table B.21: Single-stream sensors – Units, filter and
scheduled calibration configuration (continued)