Information INF13/114–EN

Navigator 550

Hydrazine wet-section
Spares replacement procedures

1 Safety

2 Tools required

Pozidrive screwdriver

Slot-head terminal screwdriver

Anti-static strap.

3 Replacement procedures

Warning.

These procedures must be carried out by a trained technician.

CHEMICAL

Ensure personal protective equipment (PPE) such as

gloves

and

eye

protection

are

worn

during

any

maintenance.

Observe all health and safety procedures for handling
chemicals.

To familiarize yourself with handling precautions, dangers
and emergency procedures, always review the Material
Safety

Data

Sheets

prior

to

handling

containers,

reservoirs and delivery systems that contain chemical
reagents and standards.

Take care if cleaning any spillages and observe all relevant
safety instructions. Wipe up any spillages using clean
water.

Perform general cleaning of the wet-section using a damp
cloth only – mild detergent can be used as a cleaning aid.
Do not use Acetone or any organic solvents.

ELECTRICAL

Isolate all high voltage supplies to the transmitter before
performing replacement procedures.

The wet-section is vulnerable to electrostatic damage.
Wear an anti-static strap or dismantle the wet-section on
an anti-static workbench.

Ensure all electrical connections are kept dry at all times.

GENERAL

Shut off the external sample supply to the wet-section
before performing replacement procedures. When a
procedure is complete, restore power to the transmitter,
sample to the wet-section at the correct flow rate and, if
necessary, calibrate the wet-section – refer to the
wet-section Operating Instructions (OI/AHM550-EN) for
calibration instructions.

Dispose of the old components in accordance with the
guidelines

contained

in

the

Operating

instructions

(OI/AHM550-EN).

Fig. 3.1 Low level dissolved oxygen wet-section – replacement procedures

Modbus cable assembly –

see Section 3.1, page 2

PCB housing seals,

PCB cover and PCB –

see Section 3.1, page 2

Flowmeter assembly

and flowmeter cable –

see Section 3.2, page 2

Constant-head

assembly, pressure

switch assembly and

pressure switch cable –

see Section 3.4, page 3

Solenoid valve assembly

and solenoid valve cable –

see Section 3.4, page 3