FEP630, FEH630
ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
111
Error no. / Range
Text on the LCD display
Cause
Remedy
F088.012 / electronics
FEB communication error.
EMC disturbance. Call Service.
EMC interference on the signal cable.
Wrong signal cable.
Check signal cables and connection
Please contact the service department.
F086.018 / Electronics
Curr.Out 31 / 32 com error.
Defective Board. EMC
disturbance. Call Service.
Broken motherboard hardware.
EMC interference
Please contact the service department.
F084.010 / electronics
NV data defect.
Data storage irreparable. Call
Service.
Data in SensorMemory corrupt.
Please contact the service department.
F082.013 / Electronics
Incompatible Frontend Board.
Frontend not fit to
Motherboard. Call Service.
Wrong frontend board or motherboard.
Please contact the service department.
F081.025 / electronics
MB voltages outside range.
Defective Motherboard HW. Call
Service.
Broken motherboard hardware.
Replace motherboard.
Please contact the service department.
Function check
Error no. / Range
Text on the LCD display
Cause
Remedy
C078.003 / Config.
Flowrate to zero.
Check digital in terminals.
The Digital Input option card is configured to
trigger the ’set flowrate to zero’ option and this
event.
Check terminals of digital input and
configuration.
C076.005 / Config.
All totalizer stopp.
Check digital in terminals.
The Digital Input option card is configured to
trigger the ‘All Totalizer stop’ option and this
event.
Check terminals of digital input and
configuration.
C074.006 / Config.
Totalizer reset.
Reset of one or more Totalizers.
The Digital Input option card is configured to
trigger the Reset Totalizer option and this event.
Check terminals of digital input and
configuration.
C072.002 / Config.
Simulation is on.
Simulated values.
Switch off Simulation Mode.
The simulation of a process value or an output is
active.
Switch off simulation mode.
C070.026 / Config.
An alarm is simulated.
Switch off alarm simulation.
The simulation of an alarm is active.
Set alarm simulation to ‘Off’.