TBN580 TRANSMITTER
CALIBRATION
K
-
9
The value displayed is the temperature of the pro-
cess (interior of sensor). Default is 25
°
C.
NOTE:
If a new temperature value is entered, it is
used only for pH calibration. This change does not
affect the calibration on the temperature sensor.
Default values 4.01 and 7.00 pH. Enter low pH value
and press
. Enter the high pH value. If
mV
SOURCE
, default is 0.00 and 14.00 pH at 25
°
C.
Displays low pH calibration value, and the uncali-
brated pH value as determined from the input
source.
Displays high pH calibration value, and the uncali-
brated pH value as determined from the input
source.
Value from the last calibration. If
NEW SENSOR
effi-
ciency will be 98%.
This value is recalculated during each calibration.
This value shows the offset change over time for a
particular sensor. Value resets to 0.00 pH if installing
a new sensor.
Key
Display
Comments
E N T E R
CHANGE pH CAL
TEMPERATURE?
nnn
°
UNITS
E N T E R
ENTER LO pH CAL
nn.nn pH
ENTER HI pH CAL
nn.nn pH
ENTER
E N T E R
APPLY nn.nn pH
nn.nn pH
HIT ENTER
WHEN STABLE
E N T E R
APPLY nn.nn pH
nn.nn pH
HIT ENTER
WHEN STABLE
E N T E R
CALCULATED
EFFICIENCY
nnn.n %
HIT ENTER
E N T E R
CALCULATED
SENSOR OFFSET
nnn pH
HIT ENTER
E N T E R
CUMULATIVE
OFFSET CHANGE
nnn pH
HIT ENTER
E N T E R
ID TAGNAME
READY