TB82 TRANSMITTER
CALIBRATION
N
-
9
Edit Calibration
This procedure describes how to edit a calibration.
Key
Display
Comments
This operation will cause a change in output not cor-
responding to the input. Selecting
YES
will hold the
output during calibration.
Select
YES
.
Select
EDIT CAL
.
Enter the sensor slope value and press
.
Valid values are 40% to 150%.
Enter a sensor offset value. Valid values are -1,000
to +1,000 mV.
Enter a temperature slope value and press
.
Valid range is 0.2 to 1.5.
Enter a temperature offset value. Valid range is -40°
to 40°C.
If any warnings or messages exist pertaining to this
device, the terminal will list them.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
E N T E R
nnnn CAL
TEMP. CAL
→
EDIT CAL
RESET CAL
E N T E R
PROCESS
SLOPE :
nnn.nnnn
OFFSET:
nnnn.nnnn
ENTER
E N T E R
TEMPERATURE
SLOPE :
nn.nnnn
OFFSET:
nnn.nnnn
ENTER
E N T E R
ID TAGNAME
READY