600T & 600T EN PRESSURE TRANSMITTER
CREATE/MODIFY CONFIGURATION
Q - 16
Input lower range value using the number keys, then
press
. Input the upper range value, then
press .
Select the 600T EN
TEMPERATURE UNITS.
Use arrow key to select option, then press
.
Select
YES
.
Key
Display
Comments
E N T E R
LOWER RANGE VAL
nn.nn UNITS
UPPER RANGE VAL
nn.nn UNITS
ENTER
ENTER
E N T E R
TEMPERATURE UNITS
→ °
C
°
F
°
R
°
K
ENTER
E N T E R
STORE THIS
CONFIGURATION?
NO
→
YES
E N T E R
ID TAGNAME
READY