EBTH TEMPERATURE TRANSMITTER
CALIBRATION
B
-
5
CALIBRATION
This section details the EBTH temperature transmitter calibration
functions of the STT04 terminal. There are two types of calibration
functions:
•
Bench calibration.
•
D-to-A adjust.
NOTES:
1.
When calibrating, a field device must be connected to the
STT04 terminal.
2.
To change calibration or configuration parameters of a field
device that is connected to an IMFBS01 module, the transmitter
must be taken off-line. This is done at the INFI 90 OPEN console.
Refer to Figure
for an overview of the calibration functions.
Select
YES
.
Configuration ID tag name just configured will be in
the upper left corner if the configuration is saved. It
becomes the working configuration.
Key
Display
Comments
E N T E R
STORE THIS
CONFIGURATION
NO
→
YES
E N T E R
ID TAGNAME
READY
Figure B-2. Calibration Flowchart (EBTH)
A P P LY 0 m V
S O U R C E
A D JU S T H IG H
C A L IBR AT IO N
R E SIS TA N C E
B EN C H
C A LIBR ATIO N
A D JU S T LO W
C A LIBR ATIO N
C A L IB R AT IO N
C O M P LE TE
A D JU S T H IG H
C A LIBR ATIO N
A PP LY 0
Ω
A PP LY 25 0
Ω
A D JU S T LO W
C A L IBR ATIO N
R E SIS TA N C E
T 01 9 10 A
A P P LY 1 00 m V
S O U R C E
C
A
L
I
B
R
A
T
E
D -T O -A
C A L
M E TH O D
A R R O W
K EY S
M E TE R VA L U E
E N TRY
A D JU S T
O U T PU T
TO 4 m A
A D JU S T
O U TP U T
TO 2 0 m A
4 m A
C A LIBR ATIO N
E N TE R M ETE R
VA LU E 4.00 0
2 0 m A
C A L IBR AT IO N
E N TE R M ET E R
VA LU E 20 .0 00