HMI LCD display type AS
20 - EN
TTH200
CI/TTH200-EN
Change from one to two columns
6 HMI LCD display type AS
Change from one to two columns
6.1
Ex relevant specifications
6.1.1 Intrinsic Safety ATEX
Explosion protection
Approved for use in Zone 0.
Designation
II 1G Ex ia IIC T6
EC type-examination certificate PTB 05 ATEX 2079 X
6.1.2 Intrinsic Safety IECEx
Explosion protection
Approved for use in Zone 0.
Designation
Ex ia IIC T6
For further information, see certificate
6.1.3 Safety specifications for Intrinsic Safety
ATEX / IECEx
Temperature table
Permissible ambient temperature range
Temperature
class
Device category 1
use
Device category 2
use
T6
-40 … 44 °C
-40 … 56 °C
(-40 … 111.2 °F)
(-40 … 132.8 °F)
T5
-40 … 56 °C
-40 … 71 °C
(-40 … 132.8 °F)
(-40 … 159.8 °F)
T4
-40 … 60 °C
-40 … 85 °C
(-40 … 140 °F)
(-40 … 185 °F)
Protection type intrinsic safety Ex ia IIC
Supply circuit
Max. voltage
U
i
= 9 V
Short circuit current I
i
= 65.2 mA
Max. power
P
i
= 101 W
Internal inductance L
i
= 0 mH
Internal
capacitance
C
i
= 0 nF
6.1.4 Intrinsically
Safe
FM
I.S. Class I Div 1 and Div 2, Group: A, B, C, D or
I.S. Class I Zone 0 AEx ia IIC T*
Temp. Ident: T6 T
amb
56 °C, T4 T
amb
85 °C
U
i
/ V
max
= 9V, I
i
/ I
max
< 65,2 mA, P
i
= 101 mW
C
i
= 0,4 µF; L
i
= 0
Control Drawing: SAP_214 748
6.1.5 Non-Incendive
FM
N.I. Class I Div 2, Group: A, B, C, D or
Ex nL IIC T*, Class I Zone 2
Temp. Ident: T6 T
amb
60 °C, T4 T
amb
85 °C
U
i
/ V
max
= 9V, I
i
/ I
max
< 65,2 mA, P
i
= 101 mW
C
i
= 0,4 µF; L
i
= 0
Control Drawing: SAP_214 751
6.1.6 Intrinsically
Safe
CSA
I.S. Class I Div 1 and Div 2; Group: A, B, C, D or
I.S Zone 0 Ex ia IIC T*
*Temp. Ident T6 T
amb
56 C, T4 T
amb
85 °C
U
i
/ V
max
= 9V, I
i
/ I
max
< 65,2 mA; P
i
= 101 mW
C
i
< 0,4 F
µ
, L
i
= 0
Control Drawing: SAP_214 799
6.1.7 Non-Incendive
CSA
N.I. Class I Div 2, Group: A, B, C, D or
Ex nL IIC T*, Class I Zone 2
*Temp. Ident T6, T
amb
60 °C, T4 T
amb
85 °C
U
i
/ V
max
= 9V, I
i
/ I
max
< 65,2 mA, P
i
= 101 mW
C
i
< 0,4 µF, L
i
= 0
Control Drawing: SAP_214 750
Change from one to two columns