Safety
CI/TTH200-EN
TTH200
EN - 7
1.5.2 Name
plate
The name plate is located on the transmitter housing.
A00236
Automation Products GmbH
Alzenau
Germany 2008
TTH200
TTH200
U = +11...30 V, I = 4...20 mA, HART
Cfg: 1 x Pt100 3W; 0...100°C
TAG: 123456
S
a
O-Code: TTH200-E1H
00001234560000
Ser.-No: 123456789012
HW. Rev. 1.06
SW. Rev. 01.00.02
PTB 05 ATEX 2017 X
II 1 G EEx ia IIC T6
II 2(1) G EEx [ia]ib IIC T6
II 2G (1D) Ex [iaD]ib IIC T6
T1...T4 Ta. = -40 °C ... +84 °C (Zone 0) ... +85 °C (Zone1)
T5
Ta. =
T6
Ta. =
TTH200-E1H;
-40 °C ... +56 °C (Zone 0) ... +71 °C (Zone1)
-40 °C ... +44 °C (Zone 0) ... +56 °C (Zone1)
0102
HART
®
13
14
10
8
4
7
5
3
2
12
11
6
9
1
Fig. 1
1
Model name
2
Manufacturer of transmitter
3
Country and year of manufacture
4
Order code
5
Serial number
6
Technical data
7
Configuration
8 TAG
number
9
Ex mark (optional)
10
Marking of explosion proof design
(optional)
11
Temperature classes of explosion proof
design (optional)
12
Refer to product documentation
13
Software revision number/hardware
revision number
14
CE mark (EC conformity)
Important
The temperature range specified on the name plate (11) refers only to the permissible ambient
temperature range for the transmitter and not to the measuring element used in the measuring
inset.