Page 2
–10
2104349 rev. AD
2) Align the RTD head to the correct position for the conduit and wiring.
Tighten the union nut.
3) Using the supplied wiring interconnect drawings, wire the RTD probe at the
wiring block (see Figure 2
4) Following local codes, run the RTD wiring to the G4 EX unit.
Figure 2
–9 RTD probe wiring
5) Inside of the G4 EX enclosure, disconnect power from the termination
board. Remove the connection from J16.
6) Remove the connector plug from the RTD connection, J8, to begin field
wiring.
7) Using the supplied wiring instructions, make the field connections to the
connector plug. Re-insert into the J8 board connector.
8) Restore power to the unit.
2.2.7 6270 Optional Equipment Unit (OEU) installation
If the 6270 OEU enclosure is used, it may be configured to include other options,
including but not limited to, a battery pack to provide power to the G4 EX,
communication equipment and solar panel charger.
Following local codes for installation, this unit would normally be located in a DIV 2
or General Purpose area. Th
e unit may be mounted on a 2” pipe or mounted on
flat surface such as a wall.
The battery and solar panel are packed and shipped separately from the 6270
OEU. Before installation, inspect all power cables where they terminate and
connectors for damage.
Before beginning, review the procedure and the materials required for installation.
The 6270 OEU system with battery and solar panel is rated for
Class I, Division 2 locations only or can be installed in non-
hazardous locations.