Page 3
–2
2104349 rev. AD
With each selection, the user will see options that pertain only to that connection
method.
3.2.1 Connecting to a local port via a RS-232 cable
If the unit is configured to include the local RS-232 Communication port, the
connection is made using the RS-232 cable. This connects the laptop computer to
the local port on the meter.
3.2.1.1
Step-by-step instructions
1) Connect the MMI cable to the designated port on the PC (default is COM 0)
and to the local port located on the unit.
2) The user needs to open PCCU32. Upon selection, PCCU32 will open.
3) With the MMI cable connected, click the Connect icon on the PCCU32 main
toolbar. The Local Connect screen will display.
If the invalid security code screen displays, enter four zeroes
(0000) for the new code, and click the OK button. The G4 EX
should have defaulted to 0000 on start-up.
If the meter has been given a security code, the user needs to
ensure they have entered
that code in PCCU32’s System Setup
screen (see Figure 3
–1). However, if the meter has been cold
started, the security code will be reset to factory default, and
historical data will be lost.
Figure 3
–1 System Setup Screen