Steps
Do This
Edit Basal Limits.
10.
STOP BASAL DELIVERY
a.
to edit basal limits. See Section 3.1.3 for procedure
as needed.
Scroll up to highlight
b.
Increment.
Press
c.
Enter button.
Scroll to select desired increment.
d.
Press
e.
Enter button.
Notice
f.
Maximum is highlighted.
Press
g.
Enter button.
Scroll to select desired maximum.
h.
Press
i.
Enter button.
Review settings, and make changes as needed.
j.
Notice
k.
Done is highlighted
Press
l.
Enter to return to the Home screen.
RESTART BASAL DELIVERY
m.
if you have no other edits.
You have completed editing the basal limits.
Edit Temp Basal Setup.
11.
Scroll to highlight
a.
Temp.
Press
b.
Enter button.
Scroll to select
c.
Percent or Units.
Press
d.
Enter button.
Notice
e.
Done is highlighted.
Press
f.
Enter button to return to the Home screen.
You have completed editing the temp basal setup.
Steps
Do This
Edit Bolus Limits.
12.
Scroll up to highlight
a.
Increment.
Press
b.
Enter button.
Scroll to select desired increment.
c.
Press
d.
Enter button.
Notice
e.
Maximum is highlighted.
Press
f.
Enter button.
Scroll to select desired maximum.
g.
Press
h.
Enter button.
Review settings. Notice
i.
Next is highlighted.
Press
j.
Enter button.
Notice
k.
ONE-BUTTON BOLUS screen displays.
Scroll up to highlight
l.
On/Off.
Press
m.
Enter button.
Scroll to select
n.
On or Off.
Press
o.
Enter button.
If you selected
Then do this
On
Notice
1.
Increment is highlighted.
Press
2.
Enter button.
Scroll to select desired increment.
3.
(0.5 u or 1.0 u).
Press
4.
Enter button.
Review settings, and make changes as needed.
5.
Notice
6.
Done is highlighted.
Press
7.
Enter button to return to the Home screen.
Off
Notice Done is highlighted and Increment is gone from the
screen.
Press Enter button to return to the Home screen.
You have completed editing the bolus limits.
Section 6: When You’re Ready – Optional Settings
6.2 Edit Additional Options
Section 6: When You’re Ready – Optional Settings
6.2 Edit Additional Options