CELL-DYN
®
1800 OPTICON OPT-6125 Trigger-Activated Hand-Held Bar Code Scanner User’s Guide
9140394A—November 2008
1
Scanner Installation:
Materials Required:
•
OPTICON OPT-6125 Bar Code Scanner
•
Adapter Cable, 5-Pin AT to 6-Pin PS/2
•
Bar Code Scanner Holder (optional)
1. From the
MAIN MENU
press
SPECIAL PROTOCOLS
, then
DAILY
SHUTDOWN
. Wait for the sequence to complete (approximately 5
minutes).
2. Power OFF the CELL-DYN 1800 System.
A
.
Install New Bar Code Scanner
1. Disconnect the keyboard cable from the right side of the
CELL-DYN 1800 System.
OPTICON OPT-6125 Barcode Scanner with Adapter Cable
2. Plug the CELL-DYN 1800 Keyboard Cable into the Scanner Cable at the
5-pin AT connector, making sure the pins line up.
3. Plug the 6-pin PS/2 connector on the Adapter Cable into the matching
connector on the Scanner Cable, making sure the pins line up.
4. Plug the 5-pin AT connector on the Adapter Cable into the connector on the
right side of the CELL-DYN 1800 System, making sure the pins line up.
5. Attach the (optional) Bar Code Scanner Holder to the right side of the
CELL-DYN 1800 System by removing the adhesive covering and
pressing the hanger firmly to the instrument.
6. Power ON the CELL-DYN 1800 System, wait for the
INITIALIZED
message to appear in the Status Box, then press
PRIME/RUN
to bring the
CELL-DYN to
READY
.
1. Handheld
Barcode
Scanner
2. Adapter Cable
3. 6-Pin PS/2
connector
4. 5-Pin AT
connector
1
2
3
4
9140394A Book.book Page 1 Monday, November 24, 2008 12:24 PM