DIGIVAL User Manual
16
4.
Close the drawer. The
DIGIVAL automatically starts
the calibration check. Do not
open the drawer.
5.
When the calibration check
is complete, the result is
displayed for review.
•
If Calibration Check passes,
tap
OK to proceed to the
next step.
•
If it fails, the card can be
cleaned using a soft, lint-free
cloth to remove dust or
debris. Handle the calibration
check card with gloves to
prevent against fingerprints
that may effect the reading.
Also see Section 16, E.18.
•
If a printer is connected,
you can tap
Print to print a
report. The printing screen
displays.
6.
When printing has finished,
open the drawer,
remove the
Calibration Check Card and
carefully return it to the zip
lock bag to protect it from
sunlight, and replace the tray.
7.
The calibration check is
complete.