DIGIVAL User Manual
28
Settings - DIGIVAL Assay 1
Select Option
Always Read Now
Always Walk Away Mode
Back
OK
1. Tap
Walk Away Mode from the
Test Settings menu.
2. Tap the assay to configure.
3. Select the read type:
Always Read Now: All tests on
the DIGIVAL will read a test
immediately after inserting a test
device.
Always Walk Away Mode: The
DIGIVAL will always time assay
development and interpret the test
result at the read time.
7.3.6 Lot Management
Home>Settings>Test Settings>Lot Management
The
Lot Management menu allows the administrator to specify how Lot QC
is handled. Lot QC refers to the requirement of performing successful QC on
any new lot of test devices prior to running patient tests with the lot.
When
Lot QC is on, the DIGIVAL checks for prior successful QC when a
patient test is run. If successful QC has not been performed on the test device
lot, the DIGIVAL displays an error message indicating that QC tests must be
performed.
The DIGIVAL prevents the user from running a patient test with that lot until
the required QC tests have been run.
When
Lot QC is off, the DIGIVAL allows patient tests to be run without
checking for prior QC on the test device lot.
Select the desired Lot QC setting
using the touchscreen. Tap
OK to
save the new setting and return to
Test Settings menu.
Summary of Contents for DIGIVAL LFR-024
Page 1: ...LFR 024 USER MANUAL DIGIVAL...
Page 2: ......
Page 3: ...DIGIVAL User Manual DIGIVAL User Manual...
Page 6: ......
Page 77: ......
Page 78: ......
Page 79: ......