DIGIVAL User Manual
34
Uploading results to a middleware system requires that the network interface
and middleware settings be configured properly as described in sections 7.4.1
and 7.4.2.
To change the Auto Upload Results setting, tap the desired option then tap
OK
to save the new setting and return to the Connectivity Settings menu.
7.4.4 Sync Settings
Home>Settings>Connectivity>Sync
Use the Sync menu option to enable
or disable periodic connections to a
connected middleware system. If Sync
is enabled, the DIGIVAL connects
to the middleware system every 10
minutes to ensure it has the latest
configuration settings, operator list,
allowed lots list, and other settings as
configured in the middleware system.
7.4.5 Purge Test Results
Home>Settings>Connectivity>Purge Test Results
Use the Purge Test Results menu
option to configure the frequency that
the DIGIVAL purges test results after
uploading them to the configured
middleware.
Disabled (Default): The DIGIVAL will not purge any previously uploaded test
results.
Purge after upload: The DIGIVAL will purge test results on the DIGIVAL
immediately after they've been uploaded to the configured middleware.
Purge after 3 days: The DIGIVAL will purge test results on the DIGIVAL three
days after they were successfully uploaded to the configured middleware.