DIGIVAL User Manual
52
14 Care and Cleaning
Cautions:
Do not disassemble or attempt to repair the DIGIVAL or other
accessories as there is a risk of damage to the DIGIVAL. The DIGIVAL
does not contain serviceable parts and should be returned for repair.
Disassembling the DIGIVAL will also void the warranty.
Do not immerse in water or cleaning solutions.
Do not clean with soap or other solutions.
Cleaning the DIGIVAL surfaces and test device tray:
Clean the exterior DIGIVAL surfaces and the test device tray daily, and
immediately after possible patient sample contamination.
Acceptable cleaning agents include the following:
•
70% ethanol
•
70% isopropanol
•
10% bleach solution on a damp, lint free cloth
DO NOT use commercially available 10% bleach wipes for cleaning the
DIGIVAL.
Do not spray or pour solution directly onto the DIGIVAL when cleaning.
Ensure no excess liquid is used when cleaning as it may damage the DIGIVAL.
Clean surrounding bench area. Clean DIGIVAL and surrounding areas
immediately after possible patient sample contamination.