E.31 QC Lot Number Invalid
Error: QC Lot Number must be 2 to 20 characters long, contain only
alphanumeric characters and have no spaces.
Action: Tap OK and try again.
E.32 Operator Import Failed
Error: Operator Import failed to upload.
Action: Check if USB Memory Stick is connected and has an operator list in
the correct format.
E.33 Operator Export Failed
Error: Operator Export Failed.
Action: Check if USB Memory Stick is connected, formatted correctly and
has enough free memory.
E.34 Multiple Update Files Found
Error: Multiple update files were found on the USB Memory Stick.
Action: Try again with a USB Memory Stick containing an update file.
Note: For error codes not found in the user manual, please contact Abbott
Technical Support.
E.35 Synchronization Failed - Network
Error: Network can not be reached.
Action: Check cable and network configuration and try again.
If error persists, contact Abbott Technical Support.
E.36 Synchronization Failed - Server
Error: Sync failed due to server communication error
Action: Contact your administrator.
E.37 Positive Patent ID Failed - General
Error: System failed to match Patient ID entered with Patient ID information
in system
Action: Check ID and re-enter or re-scan. If error persists, check information
in Hospital Information System that manages patient records.