DIGIVAL User Manual
60
E.38 Assay Lot Number is Locked Out
Error: The lot entered has been locked out by the administrator.
Action: Tap OK to try again using a Test Device with a different lot number.
E.39 Patient ID Lockout
Error: The patient ID is unknown.
Action: Tap OK and try again.
E.40 Patient Information Lockout
Error: The information for a specific patient ID is incorrect.
Action: Tap OK and enter correct patient ID.
E.41 Input Invalid
Error: The Threshold must be greater than 0 and less than 1000 (for Rule
under QC Lot Management).
Action: Tap OK to try again.
E.42 Barcode Invalid
Error:
Barcode is not valid.
Action:
Verify that pouch barcode was scanned or manually entered correctly.
Tap
OK to try again.
E.43 Lot ID Add Failed
Error: Failure to Add Lot to the Blacklist.
Action: Tap OK to try again. If problem persists, contact Abbott Technical
Services.
E.44 Lots Import Failed
Error: Failure to import Lots from USB memory stick.
Action: Check if USB is connected, formatted correctly, and has enough free
memory.
E.45 Lots Exports Failed
Error: Failure to export Lots from USB memory stick.
Action: Check if USB is connected, formatted correctly, and has enough free
memory.