background image

Diathermy is further prohibited because it may also damage the neurostimulation system components. This

damage could result in loss of therapy, requiring additional surgery for system implantation and

replacement. Injury or damage can occur during diathermy treatment whether the neurostimulation system

is turned on or off.

Electrosurgery. 

To avoid harming the patient or damaging the neurostimulation system, do not use

monopolar electrosurgery devices on patients with implanted neurostimulation systems. If use of

electrocautery is necessary, place the neurostimulator in Surgery mode using the clinician programmer app

or the patient controller app before using an electrosurgery device.
During the implant procedure, if an electrosurgery device must be used, take the following actions:

Use bipolar electrosurgery only.


Place the neurostimulator in Surgery mode before using an electrosurgery device.


Set the electrosurgery device to the lowest possible energy setting. Output power below 80 W is

recommended for all activations.

Complete any electrosurgery before connecting the leads or extensions to the neurostimulator.


Keep the current paths from the electrosurgery device as far from the neurostimulation system as

possible.

Exit Surgery mode during intraoperative testing and after the procedure is completed.

NOTE: During intraoperative testing, Surgery mode must be turned off for the neurostimulation system

to function correctly.

Confirm that the neurostimulation system is functioning correctly during the implant procedure, before

closing the neurostimulator pocket, and after the implant procedure.

After any surgery, check the neurostimulation system for the following:

Check the neurostimulator to ensure Surgery mode has been turned off, even if Surgery mode was not

turned on at the beginning or during the procedure.

Confirm the neurostimulation system is functioning.

Implanted cardiac systems.

 Physicians need to be aware of the risk and possible interaction between a

neurostimulation system and an implanted cardiac system, such as a pacemaker or defibrillator. Electrical

pulses from a neurostimulation system may interact with the sensing operation of an implanted cardiac

system, causing the cardiac system to respond inappropriately. To minimize or prevent the implanted cardiac

system from sensing the output of the neurostimulation system, (1) maximize the distance between the

implanted systems; (2) verify that the neurostimulation system is not interfering with the functions of the

implanted cardiac system; (3) avoid programming either device in a unipolar mode (using the device’s can as

an anode) or using neurostimulation system settings that interfere with the function of the implantable

cardiac system.

Other active implanted devices.

 The neurostimulation system may interfere with the normal operation of

another active implanted device, such as a pacemaker, defibrillator, or another type of neurostimulator.

Conversely, the other active implanted device may interfere with the operation of the neurostimulation

system.

Interference with other devices. 

Some of this system’s electronic equipment, such as the programmer and

controller, can radiate radiofrequency (RF) energy that may interfere with other electronic devices, including

other active implanted devices. Avoid placing equipment components directly over other electronic devices.

To correct the effect of interference with other devices, turn off the equipment or increase the distance

between the equipment and the device being affected.

Operation of machines, equipment, and vehicles.

 Patients using therapy that generates paresthesia should

turn off stimulation before operating motorized vehicles, such as automobiles, or potentially dangerous

machinery and equipment because sudden stimulation changes may distract them from properly operating

it. However, current data shows that most patients using BurstDR™ stimulation therapy do not experience

paresthesia. For patients who do not feel paresthesia, sudden stimulation changes are less likely to occur

and distract them while operating motorized vehicles, machinery, or equipment.

Explosive and flammable gases.

 Do not use a clinician programmer or patient controller around explosive or

flammable gas fumes or vapors. This includes oxygen-enriched environments such as hyperbaric chambers.

Operating the device near gas fumes or vapors could cause them to catch fire. If gas fumes or vapors catch

fire, it could cause severe burns, injury, or death.

2

Summary of Contents for Eterna 32400

Page 1: ...Eterna Implantable Pulse Generator Model 32400 Clinician s Manual...

Page 2: ...California to cause cancer and birth defects or other reproductive harm For more information go to www P65Warnings ca gov Indicates a trademark of the Abbott group of companies Indicates a third party...

Page 3: ...xcursion Temperature Specifications 11 Product Materials 11 Generator Specifications 11 Adapters and Extensions for Compatibility with Non Abbott Leads 12 Appendix B System Components and Accessories...

Page 4: ...ii...

Page 5: ...sting of conditionally approved components refer to the MRI procedures clinician s manual for neurostimulation systems available online at medical abbott manuals For more information about MR Conditio...

Page 6: ...t with the sensing operation of an implanted cardiac system causing the cardiac system to respond inappropriately To minimize or prevent the implanted cardiac system from sensing the output of the neu...

Page 7: ...ed Patients should not be dependent on drugs and should be able to operate the neurostimulation system Infection Follow proper infection control procedures Patients should avoid charging their generat...

Page 8: ...g them on the ground Care and handling of components Use extreme care when handling system components prior to implantation Excessive heat excessive traction excessive bending excessive twisting or th...

Page 9: ...0 minutes are recommended Wireless use restrictions In some environments the use of wireless functions for example Bluetooth wireless technology may be restricted Such restrictions may apply aboard ai...

Page 10: ...etooth wireless technology NOTE This manual provides instructions for implanting the generator For instructions for using other components or more information about the neurostimulation system see the...

Page 11: ...ician s manual for the clinician programmer NOTE Before the surgical procedure set up communication between the clinician programmer and the generator while the generator is in its sterile packaging t...

Page 12: ...er Retracting the setscrew too far may cause it to come loose and fail to secure the lead or extension to the generator CAUTION To avoid sharply bending and damaging the lead or extension when perform...

Page 13: ...nd the generator in loops no smaller than 2 2 5 cm 0 98 in in diameter to provide strain relief for the lead or extension and generator connection CAUTION Do not bring the suture needle in contact wit...

Page 14: ...rt Checking the Status of the Generator Battery The generator contains a rechargeable battery The amount of time that the battery will provide active stimulation depends on the patient s stimulation s...

Page 15: ...ial Generator Titanium silicone rubber epoxy resin Pocket sizer Polybutylene terephthalate Port plug Polyether ether ketone PEEK with barium sulfate NOTE These components are not made with natural rub...

Page 16: ...onic program governs the maximum frequency 1200 number of stim sets NOTE The maximum current depends on the impedance frequency and pulse width settings Adapters and Extensions for Compatibility with...

Page 17: ...or Model Description 32400 Eterna implantable pulse generator Generator Accessories Model Description 1101 Torque wrench 12710 Port plug Charging System Model Description 16000 Charger Charging System...

Page 18: ...lead introducer 1105 Lead anchor butterfly 1106 Lead anchor long 1109 Strain relief 1112 Tunneling tool 12 in 1114 Epidural needle 14 gauge 4 in 10 cm 1116 Epidural needle 14 gauge 6 in 15 cm 1120 Tu...

Page 19: ...sessions between the generator and the patient controller or clinician programmer If the generator was stored outside of recommended storage temperatures Nominal stimulation settings for BurstDR prog...

Page 20: ...nterference received including interference that may cause undesired operation Modifications not expressly approved by the manufacturer could void the user s authority to operate the equipment under F...

Page 21: ...n issues please refer to Troubleshooting for Wireless and Coexistence Issues page 17 Wireless Security Measures The wireless signals are secured through device system design that includes the followin...

Page 22: ...onstrated safety in the MR environment within the defined conditions At a minimum address the conditions of the static magnetic field the switched gradient magnetic field and the radiofrequency fields...

Page 23: ...n the European Community European conformity affixed according to the relevant provisions of European Council Regulation 2017 745 NB 2797 and RE directive 2014 53 EU Annex II Hereby Abbott Medical dec...

Page 24: ...Table 8 Symbols and definitions Symbol Definition This equipment is certified for type certification pursuant of Article 38 24 of the Japan Radio Law 20...

Page 25: ...21...

Page 26: ......

Page 27: ......

Page 28: ...2022 09 ARTEN600144282 B 600144282...

Reviews: