* Includes children 6-17 years of age. No YSI measurements were obtained for children 4-5 years of age.
Concurrence of System and Reference (CGM vs. YSI)
The percentage of concurring glucose values (CGM vs. YSI) in each glucose
reference range is presented for each CGM range in Tables 3a and 3b and for each
YSI range in Tables 3c and 3d. For example, in the Adult subject group, when the
System glucose readings were within the 81 to 120 mg/dL range, actual blood
glucose values were between 40 and 60 mg/dL 0.2% of the time, between 61 and
80 mg/dL 9.2% of the time, between 81 and 120 mg/dL 71.1% of the time, between
121 and 160 mg/dL 19.1% of the time, between 161 and 200 mg/dL 1.0% of the
time, and between 201 and 250 mg/dL 0.1% of the time.
Table 3a: Concurrence Analysis by CGM Glucose Level (Adult; n=200)
† Levels out of System dynamic range.
Table 3b: Concurrence Analysis by CGM Glucose Level (Pediatric*; n=168)
DOC44383_rev-C 1:14pm 22-SEP-2021