Receiver Size
Height: 2.5 in. (6.3 cm). Width: 3.24 in. (8.2 cm).
Depth: 0.88 in. (2.2 cm).
Receiver Weight
3.5 oz. (99.2 grams) – including batteries.
Receiver Battery Life
Up to 60 days.
Automatic Shutoff
Built-in blood glucose meter: 2 minutes after last user
action. Receiver: 12 seconds after last user action.
Receiver Memory
• 60 days of normal use including continuous glucose
readings (stored every 10 minutes) and daily blood
glucose readings.
• Date/time will be remembered for 5 minutes after receiver
battery removal.
Calibration
Plasma equivalent.
Calibration Time
• 1st calibration: Must be performed at approximately
10 hours after a new sensor has been inserted. The
first calibration can be performed after the 10 hours.
Continuous glucose readings will NOT be reported until
the 1st calibration is performed successfully.
• 2nd calibration: Must be performed between 2 and 4 hours
after the 1st calibration or continuous glucose will not
be reported. The 2nd calibration can be performed after
4 hours; continuous glucose reporting will resume after
completing the 2nd calibration successfully.
• 3rd calibration: Must be performed between 12 and 20
hours after the 2nd calibration or continuous glucose will
not be reported. The 3rd calibration can be performed
after 20 hours; continuous glucose reporting will resume
after completing the 3rd calibration successfully.
• 4th calibration: Must be performed between 48 and 56
hours after the third calibration or continuous glucose will
not be reported. The fourth calibration can be performed
after 56 hours; continuous glucose reporting will resume
after completing the 4th calibration successfully.
Blood Glucose Test Time
(for BG mode)
Average of 7 seconds.
Blood Glucose Assay Method
(for BG mode)
Coulometric electrochemical sensor.
Blood Sample Type (for BG mode)
Whole blood, capillary.
Hematocrit (for BG mode)
15% to 65%.
18 Appendix B: Specifications
(con’t)
Customer Care: 1-866-597-5520
152
Section 18
Appendix B: Specifications
Summary of Contents for FreeStyle Navigator
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......