OVER
Performance Characteristics
Note:
Please
cons
ult your healthcare team on how to use the information in this section.
Performance of the FreeStyle Navigator® Continuous Glucose Monitoring System was evaluated in a controlled
clinical study. The study was conducted in 3 centers and included a total of 58 subjects with diabetes. Each
subject wore two FreeStyle Navigator Sensors over a 5-day period. The subjects wore one sensor on the back of
the upper arm and one on their abdomen. The FreeStyle Navigator system was calibrated with capillary finger-
stick measurements using the built-in FreeStyle Blood Glucose Meter at approximately 10, 12, 24 and 72 hours
after insertion of the sensor.All measurements were performed by a trained clinic study staff or the subject.
During the study, subjects came to the clinical center for frequent glucose samples measured once every
15 minutes on the YSI (Yellow Springs Instrument) STAT Plus™ Glucose Analyzer. YSI measurements were
performed in duplicate on venous whole blood and the FreeStyle measurements were performed in duplicate on
capillary blood from the finger. All YSI whole blood measurements were adjusted by applying a +12% correction
factor (based on a normal hematocrit value of 45%).
Users and healthcare providers should consider that performance in this study might be idealized. Participants
enrolled in the clinical study and certain conditions of the study tend to result in above average glucose control.
This, in turn, may result in the appearance that the FreeStyle Navigator Continuous Glucose Monitoring
System agrees with blood glucose levels better than it does under typical conditions. Monitors that measure
glucose in interstitial fluid often show better agreement to blood glucose levels when glucose levels are not
changing rapidly or when glucose levels are not extremely low or high. The following are some examples of why
performance of the FreeStyle Navigator Continuous Glucose Monitoring System may be idealized.
Subjects saw, on average, 15 fingerstick results per day in the clinic and 8 fingerstick results per day at home.
•
This enables subjects to control their glucose levels better.
While subjects participated in the clinic portion of the study, they were more limited in their activities than
•
someone at home. They were also provided with all their meals. Subjects who are more active, or with poor
eating habits, may create more challenging conditions for the FreeStyle Navigator Continuous Glucose
Monitoring System.
The built-in FreeStyle glucose meters used in the study were well maintained. Because the built-in FreeStyle
•
meter is used to calibrate the FreeStyle Navigator Continuous Glucose Monitoring System, performance
may be poorer if the system is not well maintained. It is important to carry out quality-control checks on
the system and code the system according to the manufacturer’s instructions to optimize performance of the
FreeStyle Navigator Continuous Glucose Monitoring System
.
18 Appendix B: Specifications
(con’t)
Section 18
153
Appendix B: Specifications
Summary of Contents for FreeStyle Navigator
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......