OVER
18 Appendix B: Specifications
(con’t)
Detection of Low Glucose
See Table 7 below for detection of low glucose. As an example, when the threshold alarm was set at 70 mg/dL
(during the day), 56 % of the low glucose events were detected by FreeStyle Navigator.
Table 7.
Low Glucose Detection
DAY
NIGHT
Low Alarm
Setting (mg/dL)
Day True
Alarms*
Day Missed
Alarms**
Day False
Alarms***
Night True
Alarms*
Night Missed
Alarms**
Night False
Alarms***
% (n/N
†
)
% (n/N)
% (n/N)
% (n/N)
% (n/N)
% (n/N)
65
46 (56/121)
54 (65/121)
19 (18/95)
80 (12/15)
20 (3/15)
41 (11/27)
70
56 (98/176)
44 (78/176)
16 (21/132)
79 (19/24)
21 (5/24)
40 (14/35)
75
59 (130/219)
41 (89/219)
9 (15/161)
72 (23/32)
28 (9/32)
37 (14/38)
85
61 (189/308)
39 (119/308)
7 (17/228)
65 (22/34)
35 (12/34)
33 (14/43)
* True Alarms are the percentage of time the glucose level was below the threshold and an alarm was activated.
** Missed Alarms are the percentage of time the glucose level was below the threshold and an alarm was not activated.
*** False Alarms are the percentage of time an alarm occurred but the glucose level was not below the threshold setting.
†
n/N is the (n)umber of observations divided by the total (N)umber.
Detection of High Glucose
See Table 8 for detection of high glucose. As an example, when the threshold alarm was set at 240 mg/dL
(during the day), 78 % of the high glucose events were detected by FreeStyle Navigator.
Table 8.
High Glucose Detection
DAY
NIGHT
High Alarm
Setting (mg/dL
Day True
Alarms*
Day Missed
Alarms**
Day False
Alarms*
Night True
Alarms**
Night Missed
Alarms***
Night False
Alarms***
% (n/N
†
)
% (n/N)
% (n/N)
% (n/N)
% (n/N)
% (n/N)
180
89 (561/630)
11 (69/630)
11 (68/628)
69 (29/42)
31 (13/42)
7 (3/44)
240
78 (295/376)
22 (81/376)
12 (47/393)
41 (12/29)
59 (17/29)
25 (7/28)
270
70 (193/274)
30 (81/274)
12 (32/265)
21 (3/14)
79 (11/14)
36 (5/14)
300
61 (117/192)
39 (75/192)
12 (20/161)
12 (1/8)
88 (7/8)
33 (1/3)
* True Alarms are the
percentage of time the glucose level was above the threshold and an alarm was activated.
** Missed Alarms are the percentage of time the glucose level was above the threshold and an alarm was not activated.
*** False Alarms are the pe
rcentage of time an alarm occurred but the glucose level was not above the threshold setting.
†
n/N is th
e (n)um
ber of observations divided by the total (N)umber
.
Section 18
161
Appendix B: Specifications
Summary of Contents for FreeStyle Navigator
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......