OVER
Steps
Do This
6.
Test the sample.
Do Not:
• Use test strips more than once. Discard used test strips.
• Scrape the blood.
• Apply blood to the flat side of the test strip.
• Apply blood to the test strip when the test strip is out of
the receiver.
• Apply blood to both edges of the test strip.
• Put blood or foreign objects into the receiver’s test
strip port.
a.
Holding the receiver at a slight angle to the sample, gently
touch the half-circle on the test strip to the sample.
Important: Your blood glucose level must be between 60 and
300 mg/dL (3.3 and 16.7 mmol/L) to be able to perform
calibration tests. If your blood glucose level is changing
rapidly, you may not be able to calibrate the system.
b.
Observe the receiver:
• The stopwatch icon does
NOT
appear within 5 seconds.
Continue adding blood sample to the
SAME
half-circle for up to 60 seconds.
• The receiver
BEEPS ONCE
(only
if Progress Tones are turned on).
Enough blood has been used.
• The stopwatch icon appears.
The receiver is measuring your blood
glucose level.
• The receiver
BEEPS TWICE
(only
if Progress Tones are turned on).
The test is done. Go to Step 7.
• Your blood glucose results appear
in the display (takes an average of
7 seconds).
• The receiver beeps three times.
The test has failed. Repeat the test.
Refer to Section 10 for error messages
and troubleshooting.
• You get an error message.
If
Then
Section 7
61
Calibrate Your System/Test Blood Glucose Manually
Proccessing 08:30A
Cancel
Summary of Contents for FreeStyle Navigator
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......