Steps
Do This
1.
Decide what to do.
2.
Reconnect the receiver to the
transmitter.
a.
Place the receiver within 10 feet of the transmitter.
b.
Decide what to do next.
• Connect your receiver and transmitter.
2
• Reconnect your receiver and transmitter.
2
• Interpret the CM screen.
3
• Interpret icons on the CM screen.
4
• Interpret icons on the Main screen.
4
• Turn the receiver backlight on.
5
If you want to
Then go to this Step
• For less than 30 minutes.
1.
Wait for the connection icon to disappear
automatically from the display to confirm
connection.
2.
Go back to Step 1 of Section 9.
• For 30 minutes or more.
1.
Press the RIGHT Option button twice to
display the Main Menu screen.
2.
Use the UP/DOWN Arrow button to highlight
System.
3.
Press the RIGHT Option button (Select) to
display the System menu with Reconnect
highlighted.
4.
Press the RIGHT Option button (Select).
5.
Wait for the receiver to beep. If it beeps:
• Twice – the system is connected. Go back to
Step 1 of Section 9.
• Three times – the system is not connected.
Make sure the battery in the transmitter
is new and ensure that the receiver and
transmitter are linked (see Section 14).
If the receiver has
Then
lost connection
Customer Care: 1-866-597-5520
82
Section 9
Daily Use
System Menu 08:30A
Reconnect
Data Upload
Status
Link
Set Date/Time
Main Select
Summary of Contents for FreeStyle Navigator
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......