What You See on the Display
What You Do
Comments
7-4
Questions?
Refer to Page 10-9 For Technical Support Instructions.
Chapter 7: Linearity Test
11.
Follow the instructions in the CVC kit
package insert. Then apply a drop of the
sample to the target area.
When sufficient sample has been applied, the
monitor beeps, displays
Sample Accepted
and automatically starts the test. For a detailed
description of the test strip target area, refer to
the Test Strips section of Chapter 1.
12.
Wait for the monitor to analyze the
sample and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
13.
Note the result. If required, scan or
manually enter the comment code, and
press
Enter
.
The monitor may be enabled to scan or
enter a 1- to 2-digit comment code. If
there is no prompt to enter a comment
code, skip down to step 14.
The display includes the level and test
numbers, date and time.
Level
indicates Level 1-5.
Test
indicates replicates 1-4 for that level.
14.
You can do one of the following options:
Press
1 – New Level
.
Press
2 – Same Level
.
Press the
Menu
button to return to the
Menu Mode
.
Press
On/Off
to turn off monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
You can press
2
to perform more tests for
the
Same Level
(return to step 8) or press
1
to perform tests for
New Level
(return
to step 7).
2.5 mmol/L
2.5 mmol/L