7.13
Impella
®
System for Use During Cardiogenic Shock
IMPELLA
®
2.5 OR IMPELLA CP
®
CATHETER COMPLETELY IN THE
AORTA or INLET AND OUTLET AREAS IN VENTRICLE AND OPEN
PRESSURE AREA IN AORTA
If the Impella
®
2.5 or Impella CP
®
Catheter is completely in the aorta or if the inlet and outlet
areas are in the ventricle and the open pressure area is in the aorta, the following alarm will
appear:
Impella Position Wrong
In this situation, the placement screen will appear as shown in Figure 7.8.
Impella Flow
3.5
System Power
MUTE
ALARM
0 . 0
DISPLAY
0 . 0
MENU
PURGE
SYSTEM
Impella CP S/N: 123456
30 - 09 - 2012 05:30
100%
CONTROL
FLOW
A U T O
L/min
4.0 Max
________________________
3.0 Min
Purge System
Purge Pressure: 600 mmHg
Purge Flow: 10.0 ml/hr
Motor
Current
8 08 /7 8 3
( 7 93)
( mA)
10 sec.
Placement
Signal
( mmHg)
92/64
( 7 5)
10 sec.
160
0
1000
0
Impella Position Wrong
1 .
R ed uce t h e P - lev el t o P - 2 un t i l i m a g i n g i s a v a i la b le.
2 .
R ep osi t i on w i t h i m a g i n g .
3 .
R et ur n t o p r ef er r ed P - lev el a n d con f i r m p osi t i on i n g w i t h i m a g i n g .
Figure 7.8 Impella CP
®
Catheter Completely in the Aorta or Inlet and Outlet Areas
in Ventricle and Open Pressure Area in Aorta (similar for Impella
®
2.5)
Actions to take:
1.
Under fluoroscopic guidance, determine the Impella
®
Catheter position.
2.
Reduce the P-level to P-2 and reposition the catheter as necessary.
7
PA
TI
EN
T M
A
N
A
G
E
M
EN
T T
O
P
IC
S
Summary of Contents for Impella 2.5
Page 4: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Page 16: ......
Page 22: ......
Page 38: ......
Page 40: ......
Page 108: ......
Page 171: ......
Page 173: ......
Page 181: ......
Page 183: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 210: ...INDEX TBD...