5.6
Instructions for Use & Clinical Reference Manual (US)
INSERT PURGE CASSETTE
1.
Open the purge cassette package.
2.
Secure the red and yellow luers to the sterile field (Impella
®
2.5 and Impella CP
®
) or
discard the Y-connector and secure the yellow luer connector on the purge tubing to
the sterile field (Impella
®
5.0 and LD).
3.
Pass the purge cassette and spike off the sterile field.
4.
Spike the fluid bag/bottle.
5
Open the purge cassette door by pressing the release on the left side of the controller.
Insert the purge cassette into the Automated Impella
®
Controller (as shown in Figure
5.4 and described in the steps that follow).
1. Snap Purge Cassette
into Compartment
3. Extend Purge Tubing and
Close Purge Cassette Door
2. Slide Purge Pressure
Transmitter into Slot
Until It Snaps into Place
Figure 5.4 Inserting Purge Cassette into Automated Impella
®
Controller
6.
Insert the purge cassette into the compartment on the front of the controller. Follow
the diagram on the inside of the purge cassette door for proper placement.
7.
Slide the purge pressure transmitter into the slot to the right of the purge cassette until
it snaps into place. The controller will automatically begin priming the purge cassette.
8.
Extend the purge tubing and close the purge cassette door. There is sufficient room
around the edges of the purge cassette door so that it will not pinch the purge tubing as
it exits.
9.
The controller automatically begins priming the purge cassette after it is inserted. The
progress bar shown in Figure 5.3 marks the progress of the purge cassette priming.
Shaded Steps
All shaded steps require
sterile technique.
Purge Solution Bottles
If the purge solution is
supplied in bottles, open
the vent on the purge fluid
spike and follow the same
procedure as if supplied in
bags.
1.
1.
Open the purge cassette package.
2.
2.
Secure the red and yellow luers to the sterile field (Impella
®
2.5 and Impella CP
®
) or
discard the Y-connector and secure the yellow luer connector on the purge tubing to
the sterile field (Impella
®
5.0 and LD).
3.
3.
Pass the purge cassette and spike off the sterile field.
Connect Purge Pressure
Transmitter Within 10
Seconds
An alarm will appear if the
purge pressure transmitter is
not snapped into place within
10 seconds of inserting the
purge cassette.
Close Purge Cassette
Door
Once the purge cassette is
installed, be sure to close
the purge cassette door to
prevent the purge cassette
from being dislodged
accidentally.
Summary of Contents for Impella 2.5
Page 4: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Page 16: ......
Page 22: ......
Page 38: ......
Page 40: ......
Page 108: ......
Page 171: ......
Page 173: ......
Page 181: ......
Page 183: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 210: ...INDEX TBD...