Calibration Mode
1. While the scale is in Weighing Mode and the weighing unit is in gram, press
button quickly
multiple times until LED display shows [
CAL
] and then [
500
].
2. Here is an example of using 500g for calibration. Place a 500g standard weight on scale and
wait for the scale to execute calibration procedure.
3. The LED display shows [
-End-
] when the calibration procedure is completed. You may
unload the calibration weight when completed.
* Note: For calibration, we recommend 500g or 1000g standard weight. If you always use the
chemex to brew your coffee, please use 1000g standard weight. Otherwise, use 500g standard
weight.
Creep Compensation Mode
Operation Steps (Before firmware 1.62):
1. While the scale is in Calibration Mode and the LED display shows [
CAL
] and then [
500
],
press
button quickly multiple times until the LED display shows [
CAL2
] and then [
2000
].
2. Place a 2000g calibration weight on scale and wait for the scale to execute creep compensation
procedure.
3. The LED display shows [
-End-
] when the creep compensation procedure is completed. You
may unload the calibration weight when completed.
Operation Steps (After firmware 1.62):
1. The creep compensation mode is disabled by default after firmware 1.62.
2. If you want to enter the creep compensation mode, please input password '2468'.
Appendix A. Smart-Calibration 20