Appendix A. Smart-Calibration
If you decide to calibrate your scale, it will only be as good as the accuracy of the calibration
weights you use.
*Note : If you use weights other than acaia Weights, we recommend standard weights with class F
1
or F
2
.
Calibration Mode
1. While the scale is in Weighing Mode and the weighing unit is in gram, press
button quickly
multiple times until LED display shows [
CAL
] and then [
500
.
00
].
2. Place a acaia 100 weights on scale and wait for the scale to execute calibration procedure.
3. The LED display shows [
_End_
] when the calibration procedure is completed. You may
unload the calibration weight when completed.
Note:
1. If your Weighing Mode is disabled, please refer to
Chapter 4: Weighing Mode
Setting
to turn
on.
2. If you use weights other than acaia Weights, we recommend standard weights with 100g*n. For
example, 500g or 1000g.
Appendix A. Smart-Calibration 25