background image

www.acclima.com   DataSnap/SnapView

 

User Manual

 

11 

 

Figure 9 

 Configuration 

 
For a DataSnap to automatically take sensor readings or allow a user to read sensors manually, it must be configured. The most 
important step is to tell the DataSnap what sensors to read. This requires that the sensors be connected to the DataSnap during the 
configuration process. It is also important to provide a name for the DataSnap and to define how frequently it will take automatic 
readings. DataSnap configuration settings are stored on the DataSnap and are retained even when power is removed.  The 
configuration wizard is divided into four steps: 

 

Connect

 

 

System Configuration

 

 

Sensors

 

 

Validation

 

 

Connecting to a DataSnap 

The initial configuration screen shows all DataSnaps that are currently connected to the computer. See figure 13. To begin 

configuring a DataSnap, click the ‘Connect’ button next to the desired DataSnap.  The name of the

 connected DataSnap will appear 

on left side of the screen and the ‘Next’ button is enabled.  Use the ‘Next’ and ‘Previous’ buttons to move through the confi

guration 

steps. All configuration operations will be performed on the connected DataSnap.  
 
Click t

he ‘Refresh List” button to scan the computer for DataSnaps and update the connection list.

 

 

System Configuration Settings 

The system configuration screen allows modification of several DataSnap settings and operations. They are: 

 

Name

 

 This is the descriptive name for the DataSnap. This is how the DataSnap will be identified throughout the SnapView 

software. The DataSnap’s unique system id is also shown but it cannot be changed. 

 

 

Interval 

-

 

This is the number of minutes between automatic sensor readings. Figure 14 shows a 

DataSnap named ‘Lab 

Experiment’

 that is configured to take a new set of readings every 20 minutes. This interval is measured from the beginning 

of the previous cycle to the beginning of the subsequent cycle. For example, if the first automatic reading cycle begins at 
10am and takes one minute to complete. The second automatic reading cycle will begin at 10:20am.  

 

System Time 

 

This sets the system time on the DataSnap. The DataSnap’s current time will be displayed. Click the 

‘Synchronize’ b

utton to set the time on the DataSnap to the same time as the connected PC. See figure 15. 

 

Advanced Options

 

 

Click on ‘Advanced Options’ to display additional operations.

 

o

 

Purge Logs 

 This will delete all logs in the memory of the DataSnap. It will not remove any logs stored on the PC. 

Summary of Contents for ACC-AGR-D01

Page 1: ...DataSnap User Manual SDI 12 Data Recorder Part Number ACC AGR D01...

Page 2: ...Up a New System 7 Selecting Acclima Moisture Sensor Locations 8 Installing Acclima Moisture Sensors 8 Wiring Sensors to the SDI 12 port 8 Using Acclima SnapView Software 10 Configuration 10 Automatic...

Page 3: ...aSnap 2 pluggable terminal blocks for the SDI 12 connection USB cable type A B AC power adapter 12V DC battery power pack Installation CD with SnapView software and driver Supported Sensors The Acclim...

Page 4: ...can be unplugged from the DataSnap while wiring sensors together Multiple sensors can be wired in parallel to the same SDI 12 port External Power Port This connection is provided should the sensor ne...

Page 5: ...via a USB cable Windows will automatically detect the new Data Snap hardware and will attempt to find the appropriate driver This driver installation process will only occur once per Data Snap Each n...

Page 6: ...because the driver was installed to this location when the SnapView software was installed Figure 3 Once the driver is found and installed the Data Snap will be assigned to a port and will be availabl...

Page 7: ...aSnap Each newly added sensor is configured with a set of default properties that can be changed at any time This is addressed later in this document Add non Acclima Sensors Each non Acclima sensor mu...

Page 8: ...ort soil moisture as a percentage of the total volume measured This means that the measured volume of soil should be free of foreign objects or variations that may alter the measurement made by the se...

Page 9: ...res to a separate single wire and then connecting the single new wire to the terminal block in the appropriate position Follow the same procedure to connect the white and blue wires The wiring should...

Page 10: ...top of the screen This will display the Configuration wizard The SDI 12 protocol requires that a sensor be given a specific address All communication with a sensor is based on this address When a sen...

Page 11: ...ormed on the connected DataSnap Click the Refresh List button to scan the computer for DataSnaps and update the connection list System Configuration Settings The system configuration screen allows mod...

Page 12: ...It is not limited to Acclima sensors The DataSnap must be configured to know which sensors to read This is done by adding sensors in the Sensors screen The current list of configured sensors is displ...

Page 13: ...e can be changed at any time The DataSnap does not support duplicate serial numbers If there are one or more sensors with the same serial number the sensor that was configured last will be used Modify...

Page 14: ...en Hovering the mouse over the icon displays the number of logs that still need to be retrieved from the DataSnap as shown in figure 18 If one or more DataSnaps are connected via a USB port the SnapVi...

Page 15: ...p s name To expand collapse the list of sensors click the gray circle or click the DataSnap s name Individual sensors can be selected deselected by clicking the checkmark next to the sensor name Figur...

Page 16: ...to be easily pasted into any Windows application View Graphs Sensor readings for all sensors can be viewed visually through the Graph screen To see the Graph screen click on the Graph button at the t...

Page 17: ...right will change the displayed start date in the main graph to a later date The right most range indicator behaves similarly and effects the end date on the main graph The selected date range is dis...

Page 18: ...e viewed by selecting the Errors in the list to the left Errors are displayed in a separate grid When both grids are displayed the grid size can be adjusted To do this hover the cursor over the horizo...

Page 19: ...can be selected or not selected as desired If an item is selected it will be displayed in the grid If it is not selected it will be hidden in the grid To select all items click the Select All option...

Page 20: ...Field Test Only data between 5am on January 16 2012 and 9 30pm on January 23 2012 will be exported The output file will be named myDataSnapExport csv and will be saved in the C data folder Tools The t...

Page 21: ...eric COM port on that computer This COM port should be visible in Device Manager as Acclima Virtual COM port The communications settings for this COM port are unimportant and can be ignored since the...

Page 22: ...22 2015 2017 Acclima Inc Toll Free 1 866 887 1470 Acclima Inc 1763 West Marcon Lane Ste 175 Meridian ID 83642 USA www acclima com...

Reviews: