18. Cable Routing for Mattress Pump ---
The diagrams below show a method for
securing the power cable (shown in yellow)
to the Empresa bed frame. The power supply
is drawn at the head end of the bed, and the
accessory at the foot end. Some parts have
been removed to improve clarity.
The main view in Figure 18.1 shows the
mattress platform profile sections raised; the
detail views in Figures 18.2 and 18.3 show
the profile sections lowered.
Figure 18.1
Figure 18.2
Figure 18.3
Route the power cable alongside the bed
power cable from the control box. Ensure
there is enough slack to enable the cable to
move with the frame as it raises and lowers.
See Figure 18.4
Figure 18.4
Figure 18.5
WARNING
When fitting accessories with a power
cable to the Empresa, the bed must be
disconnected from the mains power
supply.
Careful consideration needs to be given to
the cable route to ensure that their
function is not compromised by crushing
or shearing from the bed mechanism.
Fitting must be carried out by a competent
person. All bed functions must be tested
through a full cycle by a competent
person after fitting the accessory.
Cable trap zone when headrest or footrest is lowered
Cable trap zone when bed is lowered
Cable tie attachment points shown in red
To mattress pump
To power socket
1
2
3
4
5
1
2
3
1
Head End
Foot End
Head End
Foot End
5
2
3
4