218
Display Shows Possible Cause(s)
Possible Solution(s)
Invalid Bolus
Advice Times
j
The acting time value is less
than the offset time.
j
The acting time value must be set equal to or
greater than the offset time. Reset the acting time
value or revise the offset time.
Invalid Date
j
The date entered is invalid
(dates for reminders cannot
be set to occur in the past).
j
Re-enter date.
Invalid Hyper
Values
j
The hyper warning limit value
must be greater than all of
your target ranges in the time
block settings.
j
Reset the hyper warning limit or revise the target
ranges in time blocks and re-enter the hyper
warning limit.
j
Enter a hyper warning limit that is above the target
ranges of your time blocks. See Chapter 5,
Changing Meter Settings.
Invalid Hypo Values
j
The hypo warning limit value
must be less than all of your
target ranges in the time
block settings.
j
Reset the hypo warning limit or revise the target
ranges in time blocks and re-enter the hypo
warning limit.
j
Enter a hypo warning limit that is below the target
ranges of your time blocks. See Chapter 5,
Changing Meter Settings.