219
8
Display Shows Possible Cause(s)
Possible Solution(s)
Invalid Record
Time/Date
j
The time/date entered is
invalid (Add Data entries
cannot be set to occur in the
future).
j
Re-enter time/date.
Invalid Target
Range
j
The lower target range value
is above the upper target
range value.
j
Reset the target range values.
j
Enter the correct lower target range value and
upper target range value. See Chapter 5, Changing
Meter Settings.
j
The range selected conflicts
with your hyper and/or hypo
warning limit settings.
j
Reset range or revise warning limit settings and
re-enter range.
j
Enter the correct lower target range value and
upper target range value. See Chapter 5, Changing
Meter Settings.
LO bG Warning
j
Your blood glucose may be
lower than the measuring
range of the system.
j
Treat your low blood glucose as recommended by
your healthcare professional.
j
If you are experiencing any of the common
symptoms of low blood glucose, contact your
healthcare professional immediately.
Meter Battery Low
j
Battery power is low.
j
Install new batteries. See Chapter 7, Care and
Maintenance.