67
3
Out-of-Range Control Test Results
If your control test result is not inside the acceptable range, here are some things you can do to solve the
problem:
Troubleshooting Checks
Actions
1. Were either test strips or control solutions expired? Discard the test strips or control solutions if either is
past the use by date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.
2. Did you wipe the tip of the control solution bottle
before use?
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
3. Were the caps on the test strip container and the
control solution bottle always closed tightly?
Replace the test strips or control solutions if you
think either was uncapped for some time and repeat
the control test.
4. Was the test strip used immediately after it was
removed from the test strip container?
Repeat the control test with a new test strip and a
fresh drop of control solution.
5. Were the test strips and control solutions stored in
a cool, dry place?
Repeat the control test with a properly stored test
strip and control solution.
50839_06434240002.indb 67
7/2/12 3:25 PM
Summary of Contents for Aviva Expert
Page 8: ...6 50839_06434240002 indb 6 7 2 12 3 24 PM ...
Page 234: ...232 50839_06434240002 indb 232 7 2 12 3 26 PM ...
Page 244: ...242 50839_06434240002 indb 242 7 2 12 3 26 PM ...
Page 252: ...250 50839_06434240002 indb 250 7 2 12 3 26 PM ...
Page 258: ...256 W warnings list of 194 198 weight loss 52 50839_06434240002 indb 256 7 2 12 3 26 PM ...