89
4
i
NOTE
• On the right side of the graph, the maximum bolus amount indicator is a horizontal line with the
corresponding value shown above it. The purpose of this indicator is to scale the top portion of the
graph for the bolus value blue bars. The maximum bolus amount indicator values available on the
meter are 1, 5, 15, 30, and 60 U. The indicator displayed is based on the single largest bolus insulin
amount delivered for the time range chosen. For example, if the largest bolus delivered during the
chosen time range is 8 U, then the meter will scale the upper graph to be between 0 and 15 U.
• On the right side of the graph, the maximum carbs amount indicator is a horizontal line with the
corresponding value shown below it. The purpose of this indicator is to scale the bottom portion of
the graph for the carb value dark orange bars. The maximum carbs amount indicator values available
on the meter are 30, 60, 120, 180, and 240 g, or the equivalent scale for BE, KE, or CC. The indicator
displayed is based on the single largest carbs amount for the time range chosen. For example, if the
largest carbs amount for the chosen time range is 86 g, then the meter will scale the lower graph to
be between 0 and 120 g.
• Select Table to change the display to the table view. The time range remains the same.
50839_06434240002.indb 89
7/2/12 3:25 PM
Summary of Contents for Aviva Expert
Page 8: ...6 50839_06434240002 indb 6 7 2 12 3 24 PM ...
Page 234: ...232 50839_06434240002 indb 232 7 2 12 3 26 PM ...
Page 244: ...242 50839_06434240002 indb 242 7 2 12 3 26 PM ...
Page 252: ...250 50839_06434240002 indb 250 7 2 12 3 26 PM ...
Page 258: ...256 W warnings list of 194 198 weight loss 52 50839_06434240002 indb 256 7 2 12 3 26 PM ...