72
Testing blood glucose
Farbe: P 249 – Prüfmittelnummer: 203
73
Testing blood glucose
4
b
a
Notes
The Accu-Chek Softclix Plus lancing device is intended only for
patient self-monitoring by a single person. The lancing device
must not be used on more than one patient due to the risk of
infection.
Healthcare professionals testing blood glucose in a multi-patient
setting must not use the Accu-Chek Softclix Plus lancing device
to collect blood and should therefore remove the lancing device
from the meter. The lancing device may be replaced with a blank-
ing plate (see Chapter 10.1).
Only use Accu-Chek Softclix lancets in your Accu-Chek Softclix
Plus lancing device. Using any other lancets may damage the
lancing device or impair its function. This may result in injury.
4.2.1 Undocking the lancing device from the meter
If you wish to use the lancing device undocked from the meter, detach it
as follows:
P
Push the slide button in the direction indicated by the arrow (1) and
hold it there.
P
Withdraw the lancing device in the direction indicated by the arrow
(2).