178
Symbols, error messages and troubleshooting
Farbe: P 249 – Prüfmittelnummer: 203
179
Symbols, error messages and troubleshooting
12
Symbols
Meaning
O
because you have just inserted a
partly used test strip drum,
O
because you have just opened the
drum compartment cover.
Perform a performance check (see Chap-
ter 7). If the test produces a correct result,
you may continue to use the test strips. If
the result is incorrect, you need to insert a
new drum. Following the performance
check, the bottle symbol disappears from
the display. The drum symbol continues to
be displayed in order to draw your atten-
tion to the fact that the meter cannot cal-
culate the use-by period. The symbol only
disappears when you insert a new test
strip drum. All results are saved along with
the drum symbol. If you do not perform a
performance check, the flashing bottle
symbol appears with every test until you
insert a new drum.
12.1.3 Symbols appearing in the numeric field
In the numeric field:
88.8
(mmol/L meter) or
888
(mg/dL meter) the
following symbols may appear:
Symbol
Meaning
18
In memory:
O
No results present in memory.
O
Averaging is not possible because
you have not set the time and date,
memory only contains results with-
out a time and date,
memory only contains results that
are not included in the calculation,
e.g. performance results,
no results were saved in the period
in question, e.g. all results in mem-
ory are more than 7 days old,