77
Checking your meter
Figures are examples
After flagging the result as a performance check:
12
Compare the test result with the concentration
table on the label attached to the test strip
container.
The result must be within the specified concentra-
tion range.
Make sure that you compare the test result with the
concentration data that corresponds to the relevant
control solution that you used (1 or 2) and to the unit
of measurement your meter displays (mg/dL or
mmol/L).
Accu-Chek
®
Active Control
mg/dL
mmol/L
1
37 – 67 2,1 – 3,7
2
134 – 181 7,4 – 10,0
Control solution
used
Unit of
measurement