216
The errors the meter may display are:
Error Title
Error Message
Left Soft Key Right Soft Key
e-51
Bad Strip Error
Reinsert or replace with new Strip
e-52
code key error
Turn Meter off, reinsert or replace code key
Menu
1
OK
2
e-53
Bad Test Error
Retest with new Strip
e-54
Not enough Sample
Retest with new Strip
e-55
code key expired
Turn Meter off, replace code key and Strips
Menu
1
OK
2
e-56
Sample Applied early
Retest with new Strip
e-57
electronic error
Remove batteries, wait 20 seconds, replace
batteries
e-58
Temp. error
Move Meter to correct temperature and
wait 5 minutes
Menu
3
OK
3