121
Acoustic Mode
7
Examples:
O
The result 13.8 mmol/L is announced as follows:
1 beep – pause – 3 beeps – pause – 1 very short beep – pause–
8 beeps ( – – – – · – – – – – – – – )
O
The result 4.0 mmol/L is announced as follows:
1 long beep (for 0 tens) – pause – 4 beeps – pause – 1 very short
beep – pause – 1 long beep ( — – – – – · — )
The result announcement is repeated twice. Again, two short beeps pre-
cede each repetition.
Results that are lower than 10 mg/dL or 0.6 mmol/L and are displayed
as
Lo
are announced acoustically as 0 – 0 – 0 and 0 – 0 – decimal
point– 0, respectively.
Results that are higher than 600 mg/dL or 33.3 mmol/L and are dis-
played as
Xi
are announced acoustically as 9 – 9 – 9 and 9 – 9 – deci-
mal point – 9, respectively.